|     Written By: Dan Alcox     |     Date Posted: 03-31-2016     |

Key Takeaways

  • Capture the analog envelope
  • Use a picoscope
  • Screen shot the analog envelope

The analog envelope output is the log compressed voltage output of the acoustic return signal, which provides users with the ability to see the internal workings of the sensor. This output allows users the ability to easily verify the overall operation of their system application. Advanced users may chose to read the analog envelope and process the data with their own digital converter. This output can also provide Maxbotix Inc., with the data needed to assist with troubleshooting or to provide a semi-custom firmware solution for an application.

Troubleshooting with the Analog Envelope

The most common example of using the analog envelope is to capture interference sources, such as physical objects cluttering the acoustic path to the target. Another example of use is to see if there is another acoustic noise source nearby.

Analog envelope Clutter at 3.3 volts
Figure 1. Analog Envelope Output

Picoscope Minigrabbers
Figure 2. Picoscope & Minigrabbers

Setup Instructions

A Picoscope is a USB connected software oscilloscope displays electrical signals as they change over time, which works great for reading the analog envelope.

  1. Connect sensor ground, and V+ to a power supply set at 5V.
  2. Using probe from Picoscope, connect the main clip lead of probe to pin 2 for the analog envelope output.
  3. Connect the small clip lead from probe to the ground pin 7 of the sensor.
  4. Open the picoscope software on your computer.

How to capture the analog envelope

1) The transmit burst is the highest amplitude signal and should be used to trigger the Picoscope. Set the scope to trigger/sync on the transmit burst which is typically ~2.9V with a 5V input. (Although auto triggering is normally used for waveform setup, to capture a waveform, the triggering should be set to “Normal” or “Repeat” to assure the trigger occurs at the correct location on the waveform.)

Transmit Burst
Figure 3. Transmit Burst

Trigger Controls
Figure 4. Trigger Controls

2) Set the vertical control to maximize the transmit burst, target, and clutter information. When powered with 3V to 5V you can typically use 1mS per division to capture the waveform data.

voltage and division settings
Figure 5. Voltage and Division Settings

3) Set the horizontal controls to capture all of the information of interest. At room temperature the speed of sound is 58uS per cm, so if you need to capture objects out to 1 meter, then (100cm * 58uS) you must capture data for 580uS or 5.8mS. This corresponds to setting the horizontal control to 1mS per division and will allow data capture to 10mS on most oscilloscopes and should work for up-close measurements of less than 1.5 meters. For longer measurements, typically a 10mS per division will suffice.

Picoscope Settings
Figure 6. Picoscope Settings

4) Now you can capture a clear screenshot of the analog envelope output. Label your screenshot with the settings you used.

Figure 7. Analog Envelope Screenshot

For users who need custom firmware matched to their application, this output can provide MaxBotix Inc., with the data to help aid in the design of the firmware needed. In bin measurements the analog envelope can be used to capture the output data for the bin level. A user can capture the analog envelope when the bin is empty (0%), 10%, 20% etc., up to 100% full. These images then provide MaxBotix with the data required to match the firmware to the given application and incorporate the needed margins for robust operation.

Compatible Part Numbers

The following list of part numbers have the Analog Envelope output available:


The analog envelope can be a valuable output when wanting to range multiple targets, or when in an area with noise issues. Using a picoscope or an oscilloscope will be a useful tool to display the observed waveform information, and a good screenshot of the analog envelope can be used for troubleshooting on your own or sent to our technical support staff to help us trouble shoot the problem for you.

Contact our technical support team if you need any additional help or have any questions about using the Analog Envelope or other sensor outputs used by our sensors. We are here to help you succeed.

Author: Kathy Kostal  Date: 08-31-2016
Inc 5000 Maxbotix Logo Inc. Magazine Unveils 35th Annual List of America's Fastest Growing Private Companies–the Inc. 5000. MaxBotix Inc., Ranks No. 1752 on the 2016 Inc. 5000 with Three–Year Sales Growth of 213%.
Click here for full article.
Author: Cody Carlson  Date: 08-02-2016
Raspberry Pi 2 board MaxSonar sensors offer a variety of outputs including TTL serial data. This tutorial guides you through the process of setting up your Raspberry Pi 3 with a MaxBotix sensor. Click here for full article.
Author: Scott Wielenberg  Date: 07-26-2016
Full Horn Housing MaxBotix offers an expanded range of packaging options for many of our sensors. Each option provides unique benefits to certain mounting integrations. This article provides a brief overview of each option.
Click here for full article.
Author: Scott Wielenberg  Date: 07-18-2016
Sensor is Tested When providing support, our technical support team may determine that further testing at our facility is the best way to help resolve the issue that you are facing. At this point, they will start the Return Merchandise Authorization (RMA) process. This article will explain what you can expect as your ultrasonic sensor travels through our RMA process.
Click here for full article.
Author: Scott Wielenberg  Date: 07-11-2016
Typical Wall Pipe Many customers have requested the option to mount an ultrasonic sensor in a pipe. During the testing and development cycle, we discovered a number of considerations and requirements that must be met for the application to be successful. When all of these are met, a user may be able to achieve the desired level of success for measuring the liquid level inside of a pipe.
Click here for full article.
Author: Jenney Grover  Date: 06-28-2016
one of the options for outputs On April 19th, we welcomed our supporters to join us for the Grand Opening of the Build Out. Bob and Nita Gross gave a tour of the build out and their vision for the space. We continue to be in awe of the support from our community, our employees, our distributors, and our customers. Thank you for the many years of support, and we look forward to serving you in the years to come.
Click here for full article.
Signup for notification of our exciting new products and periodic new letters. We are excited to provide the latest information from MaxBotix Inc.